Also known as: screen height linearity, amplitude linearity, vertical linearity ultrasonic, vertical linearity in UT, vertical linearity echo.
Why amplitude is the one that decides
Ultrasonic acceptance criteria are written in amplitude. An indication is evaluated when it exceeds a reference level. A flaw is sized by traversing until the echo drops 6 dB or 20 dB. A weld is rejected when an indication exceeds a percentage of a distance amplitude correction curve. Every one of those is a comparison of echo heights, which makes the linearity of the amplitude display the property that sits underneath accept and reject decisions.
An instrument whose screen response compresses at the top of the graticule will under-report a large reflector. One that exaggerates at the bottom will produce indications that are evaluated but should not have been. Neither fault announces itself during a routine two-point calibration, because calibration sets one echo to one height and never asks what happens at the others.
The two-echo method
The check in AS 2083 is designed so no signal generator is needed. It uses two echoes whose true amplitude ratio is fixed by physics and cannot change during the test.
- Choose a centre echo h(n) and the adjacent echo h(n+1).
- Adjust the gain so h(n) sits at 10% of full screen height. Read h(n+1).
- Add gain to bring h(n) to 20% of full screen height. Read h(n+1) again.
- Repeat at 30%, 40%, 50% and so on up to 100% of full screen height.
- Calculate the ratio h(n)/h(n+1) at every step and compare each to the mean.
Because the two reflections come from the same reflector geometry, their real ratio is constant. Any variation in the measured ratio as the pair is walked up the screen is instrument error, and plotting the ratio against h(n) with tolerance bands shows immediately whether the error is a gradual compression, a step, or a single misread point.
The ASME criterion
ASME BPVC Section V, Article 5 states the requirement as a pass or fail rather than a plot. Under Mandatory Appendix I, two reflectors are positioned to give a 2 to 1 amplitude relationship, the larger set to 80% of full screen height and the smaller to 40%. The larger indication is then stepped by the gain control from 100% down to 20% of full screen height in 10% increments, and at every step the smaller indication must read half the larger within 5% of full screen height. Mandatory Appendix II then checks the gain control itself, confirming that a stated decibel change produces the change in indication height it should.
Article 4, which governs weld examination, points back to these checks and sets the interval: at the beginning of each period of extended use, or every three months, whichever is less.
Producing the evidence later
The check is worth nothing to an assessor if the record cannot be found. The plot or worksheet belongs with the instrument, next to its calibration certificate and its next due date, so any report generated with that instrument can be traced to proof it was verified. The free linearity tool on this site produces a chart that can be copied or downloaded for exactly that purpose, and in AMPS the result is filed on the equipment page where the flag system warns a month before the next check falls due.